FEATURES



  • Optimized for micro-volume applications



  • 10 valuable shots form a 100 μl stock solution


  • Dead time : 1 ms


  • Minimum injection volume : 3 μl


  • 2 independently controlled syringes


  • User-selectable mixing ratio 1:1 to 1:9


  • Temperature range : 0 ℃ to + 50 ℃


Drive mechanism: independent stepping motors
Number of syringes
2
Syringe type
Glass, 500 µl
Number of mixers
1
Mixer type
Micro Berger Ball
Stop mechanism
Electrovalve


Sample Consumption  
Mixing ratio
Fully variable from 1:1 to 1:9
Maximum flow rate
1.8 ml/s
Priming volume
24 µl
Minium volume per shot (total)
24 µl
Minimum injection volume
3 µl


Observation head
Number of detection windows
3
Cuvette light path
0.8 mm
Cuvette volume
2-3 µl
Minimum dead time
1 ms
Material
PEEK, FFKM (full solvent compatibility)


Synchronization with detection
Trigger
fully programmable; 5V TTL trigger in/ out available
Optical coupling
fiber optics or direct attachment


Temperature control
Temperature range
0 to +50 °C
Temperature control
Water bath circulator (optional)
Temperature probe
Optional PT100 allows direct reading of temperature in Bio-Kine


General
Dimensions
200 x140 x 640 mm
Weight
9 kg
Communication
USB
Power
110 V - 220 V