MRS-HR

Regenerative Battery Test System :

Up to 1000A (2400A), Up to 6V (80V), Up to 14kW (36kW) with energy recovery

Maker

BaSyTec (Germany)

Features

• 고전류 테스트 시스템

• 최대 1,000A (병렬 모드 시 2,400A)

• With energy recovery

Applications

• 2nd battery

Specifications


80V variant
Charge voltage
3-80V
Discharge voltage
3-80V
Voltage accuracy /
resolution
16mV / 24Bit
Channel
configurations
70A / 1.5kW
140A / 3kW
210A / 4.5kW
280A / 7kW
360A / 9kW
500A / 12kW
570A / 14kW
Efficiency charge
Up to 90%
Efficiency discharge
Up to 87%
4-wire measurement
Yes
Current accuracy /
resolution / ripple
0.05%FS / 24Bit / < 0.3% FS
Time resolution /
current rise time
1µs / < 1ms
Parallel operation
Yes, up to 4 channels
Current ranges
Optional, up to 3,
automatic range switching
Sampling rate
Output change rate
Min. 1ms       but not more than 1kHz continually per system
Standard inputs
Voltage, current,
2 Pt100/4W, NTC
Control modes
Current, voltage, power,
resistance and combinations
Cooling
Air cooled, temperature
regulated fans
Power supply
400V / 3 phases, N, GND,
power factor > 0.99
Noise
50 - 80 dB (A)
Other options
CAN interface
BSD (Battery Safety Device)
CMU cell voltage inputs
CMU Type-K, Pt100/4W or NTC temperature inputs
SSMS interface
Digital-IO or relay outputs
Software
BaSyTest
Interface
Ethernet

6V variant
Charge voltage
0-6V
Discharge voltage
1-6V
Voltage accuracy /
resolution
1mV / 24Bit
Channel
configurations
100A
200A
300A
...
600A
800A
1000A
Efficiency charge
Up to 80%
Efficiency discharge
Up to 70%
4-wire measurement
Yes
Current accuracy /
resolution / ripple
0.05%FS / 24Bit / < 0.3% FS
Time resolution /
current rise time
1µs / < 1ms
Parallel operation
Yes, up to 4 channels
Current ranges
Optional, up to 3,
automatic range switching
Sampling rate
Output change rate
Min. 1ms       but not more than 1kHz continually per system
Standard inputs
Voltage, current,
2 Pt100/4W, NTC
Control modes
Current, voltage, power,
resistance and combinations
Cooling
Air cooled, temperature
regulated fans
Power supply
400V / 3 phases, N, GND,
power factor > 0.99
Noise
50 - 80 dB (A)
Other options
CAN interface
BSD (Battery Safety Device)
CMU cell voltage inputs
CMU Type-K, Pt100/4W or NTC temperature inputs
SSMS interface
Digital-IO or relay outputs
Software
BaSyTest
Interface
Ethernet

MRS-HR


Regenerative Battery Test System : Up to 1000A (2400A), Up to 6V (80V), Up to 14kW (36kW) with energy recovery

Maker:

BaSyTec (Germany)

FEATURES


• 고전류 테스트 시스템

• 최대 1,000A (병렬 모드 시 2,400A)

• With energy recovery


APPLICATIONS


• 2nd battery

SPECIFICATIONS



80V variant6V variant
Charge voltage3-80V0-6V
Discharge voltage3-80V1-6V
Voltage accuracy / resolution16mV / 24Bit1mV / 24Bit
Channel configurations70A / 1.5kW
140A / 3kW
210A / 4.5kW
280A / 7kW
360A / 9kW
500A / 12kW
570A / 14kW
100A
200A
300A
...
600A
800A
1000A
Efficiency chargeUp to 90%Up to 80%
Efficiency dischargeUp to 87%Up to 70%
4-wire measurementYes
Current accuracy / resolution / ripple0.05%FS / 24Bit / < 0.3% FS
Time resolution / current rise time1µs / < 1ms
Parallel operationYes, up to 4 channels
Current rangesOptional, up to 3, automatic range switching
Sampling rate
Output change rate
Min. 1ms      but not more than 1kHz continually per system
Standard inputsVoltage, current, 2 Pt100/4W, NTC
Control modesCurrent, voltage, power, resistance and combinations
CoolingAir cooled, temperature regulated fans
Power supply400V / 3 phases, N, GND, power factor > 0.99
Noise50 - 80 dB (A)
Other optionsCAN interface
BSD (Battery Safety Device)
CMU cell voltage inputs
CMU Type-K, Pt100/4W or NTC temperature inputs
SSMS interface
Digital-IO or relay outputs
SoftwareBaSyTest
InterfaceEthernet