FEATURES


  • Wavelength range : 163 to 950 nm (up to 1250nm option)


  • Wavelength accuracy : ± 0.1 nm (Over full range)


  • CD range : ± 7500 mdeg (resolution 0.0001 mdeg )


  • Standard detection modes :

    CD/Absorbance/HV, Fluorescence, FD/CD,

    Fluorescence anisotropy and Polarization, LD, HPLC-CD


  • Optional modes : NIR-CD, ORD, DR-CD,

    Stopped-flow and titration, Emission fluorescence



     SPECIFICATIONS


Light source
Extremely quiet 150 W Xe and Xe (Hg) air-cooled,
tungsten available in option (air-cooled)
Monochromator
Tunable chromatic light source coupled to double grating (patented)
Wavelength range
163-950 nm (standard), 163-1250 nm (with optional detector)
Nitrogen gas purge
Only for scans < 195 nm ( no risk of damaging optics), 
high efficiency N2 purge optimized for light source,   
optical bench and sample compartment
Wavelength accuracy
±0.1 nm from 163 to 1,250 nm
Wavelength precision
±0.05 nm from 163 to 1,250 nm
Bandwidth
0 to 16 nm on full wavelength range
Stray light
< 2 ppm at 200 nm
CD resolution
0.00001 mdeg
CD range*
±7,500 mdeg
Baseline stability
±0.007 mdeg/hour
Scanning speed
0.1 ms to 20 s per data point
Data interval
0.1 nm to 10 nm in scanning mode,   
10 µs to 20 s in kinetics mode
Scanning modes
Step scan, 
adaptive scan, 
temperature scan, 
kinetics (slow or using stopped-flow)
Rms noise**  
0.015 mdeg at 185 nm using 1 nm BW, 16 s sampling 
0.01 mdeg at 200 nm using 1 nm BW, 16 s sampling 
0.007 mdeg at 500 nm using 1 nm BW, 16 s sampling 
Standard detection modes
CD, Absorbance, HV (standard all simultaneous) Fluorescence, FD/CD, Fluorescence anisotropy, HPLC-CD, LD
Optional detection mode
NIR-CD, ORD, DR-CD
UV measurement
Accuracy ±0.001 AU (built-in filters to remove second order)
Shutter
Built-in, software control
External input/output
4 in, 3 out (for external connections)
PC interface
Windows 7, 32 or 64 bits
Other options
Titrator (concentration and pH), emission monochromator
Dimensions
139 x 32 x 39 (cm, W x D x H)
Weight
35 kg